Installation Manual
8-92 Troubleshooting
6. Click on Dispense and then click on the Reinitialize button.
7. Does the Dispensing Head reinitialize?
Go to Step 8
Go to Step 10
8. Is the red Servo Motor status AOK LED illuminated?
Go to Step 9
Go to Power-up Reset Check in this subsection.
9. Is the Servo Amplifier status EXTERNAL DISABLE?
Go to Servo Amplifier Verification in
this subsection.
Go to 8.2, Troubleshooting the Servo Controller
10. Can the Dispensing Head be moved in X and Y axes with the Jog keys [Ctrl + Shift + Arrow] on
the Keyboard?
Go to Step 11
Stop here. Dispensing Head initialization is the root cause of Servo
Amplifier fault.
11. Does the Light Beacon green light illuminate when the Front Hatch is closed?
Go to “Light Beacon Signal Verification” in 8.1.4,
Interlock Fault Isolation Procedure
Go to Step 12
12. Is the red SRVO_PWR LOW LED on the XY Servo Interface PWA illuminated when the Front
Hatch is closed?
Go to Step 13
Go to 8.1.4, Interlock Fault Isolation Procedure
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes