Installation Manual
Troubleshooting 8-77
25. Measure VDC between Pin 1 and Pin 4 of the X-axis Home Sensor Cable (P/N 06-4550-00) at J3,
connected to the X-axis Home Sensor PWA.
26. Is the reading less than 2 VDC?
Home/Limit Switch PWA (P/N 60-0785-00) has
failed. See
Parts Replacement section.
Go to Step 27
27. Manually move the Dispensing Head to the left side of the dispensing chamber until it contacts the
X-axis Home Sensor.
28. Does the black Home Flag enter the slot in the Home Sensor without contacting the Sensor case?
Go to Step 29
Go to “PMAC Fatal Following Error Verification” in 8.2.1, PMAC Initialization
Fault Isolation Procedure
29. Loosen the two 3-mm bolts, reposition the X-axis Home Sensor, and torque the bolts to 20 in-lbs
(0.230 kg-m).
30. Manually move the Dispensing Head to the far left into the X-axis Home Sensor.
31. Does the black Home Flag enter the X-axis Home Sensor without contacting the X-axis Home
Sensor?
Go to Step 29
Go to “PMAC Fatal Following Error Verification” in 8.2.1, PMAC Initialization
Fault Isolation Procedure
No
Yes
No
Yes
No
Yes