8-52 Troubleshooting
14. Does the Dispensing Head complete the homing process with no FmNT error messages?
Go to Step 15
Go to Step 16
15. Did the Dispensing Head hit the mechanical hard stop in the X or Y-axis during the homing process?
Go to PMAC Fatal Following Error Verification in
this subsection.
Go to 8.2.5, PMAC Home Sensor Fault Isolation Procedure
16. From the FmNT Main Menu, click on the Jog icon.
17. In the Jog Window, point and click at several locations in the blue square to move the Dispensing
Head throughout the dispensing chamber.
18. Did the FmNT message “30735 The motor have gone open loop? Press OK to continue and
reinitialize the motion controller” appear?
Stop here. No PMAC fault. If fault reoccurs, contact
Asymtek Technical Support.
Go to Step 9 of PMAC Fatal Following Error Verification in this subsection.
No
Yes
No
Yes
No
Yes