Installation Manual

8-46 Troubleshooting
External Disable Check
1. Is FmNT open?
Go to Step 2
Go to Step 5
2. Close the Front Hatch.
3. Open FmNT by double clicking on the
FmNT icon.
4. Does FmNT initialize with no FmNT error messages (30735, 30416, and 30353) related to the
Motion Controller?
Go to 8.2, Troubleshooting the Servo Controller
Go to Step 10
5. With the Front Hatch closed, go to FmNT Main Window and click on the Jog icon.
6. Click on
Dispense and then on the Reinitialize button.
7. Does the dispenser reinitialize?
Go to Step 8
Go to Step 10
8. Is the red SERVO MOTOR AOK LED illuminated?
Go to Step 9
Go to Power-up Reset Check in this subsection.
9. Is the SERVO AMPLIFIER status EXTERNAL DISABLE?
Go to Servo Amplifier Verification in
this subsection.
Go to 8.2, Troubleshooting the Servo Controller
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes