Installation Manual

Troubleshooting 8-11
4. Does the FmNT message30753 The motor failed to close” appear when initializing FmNT?
Go to Step 5
Go to 8.1.5, XY Servo Amplifiers Fault Isolation Procedure
5. Does the FmNT message30796 The dispenser PMAC device initialization has failed” appear
when initializing FmNT before any other message?
Go to Step 6
Go to “PMAC Watchdog Error Verification” in 8.2.1, PMAC Initialization
Fault Isolation Procedure
6. The FmNT message “30416 Dispenser motor reference not found. Select OK to home the
dispenser should appear during initialization process.
7. Home the Dispensing Head by clicking on the
OK button in the bottom of Message Box, or press
[Enter] on the Keyboard.
8. Watch how the Dispensing Head homes (it should move upward in the Z-axis, then simultaneously
move in the X and Y-axis to the Home position in the front left corner of the dispensing chamber.
9. Does the Dispensing Head complete the homing process with no FmNT error messages?
Go to Step 10
Go to Step 11
10. Did the Dispensing Head hit the mechanical hard stop in the X or Y-axis during the homing process?
Go to “PMAC Fatal Following Error Verification” in
8.2.1, PMAC Initialization Fault Isolation
Procedure
Go to 8.2.5, PMAC Home Sensor Fault Isolation Procedure
11. From the FmNT Main Menu, click on the Jog icon.
12. In the Jog Window, point and click at several locations in the blue square to move the Dispensing
Head throughout the dispensing chamber.
No
Yes
No
Yes
No
Yes
No
Yes