Installation Manual

6-8 Calibration and Adjustment
To level the Scale:
1. Remove the Dispensing Area Front Cover and place it in a safe place. See Figure 6-4.
CAUTION! Lay the plate down on a flat surface. Do not stand it on edge or lean it against
something as this may bend the plate.
2. Gently apply pressure to opposite corners of the Scale Base to determine its all-
around stability.
> The Scale Base should not be rigid. It should compress and move up and down, but it
should not seem to be loose.
3. With the Circular Level on the clean Scale Pedestal, turn the two leveling screws on the back
of the Scale with the flat blade screwdriver, as necessary, until the bubble is centered. See
Figure 6-3.
> Turning a screw counterclockwise lowers that corner of the Scale Base.
> Turning a screw clockwise raises that corner.
4. When the Scale is level, replace the plastic cup, banjo wiper, Circular Level, and Weigh
Station Cover.
5. Reinstall the Dispensing Area Front Cover.