Installation Manual

Calibration and Adjustment 6-7
6. Locate and remove the three 1.5-mm mounting screws attaching the Circular Level inside the
well on the top of the DCM. See Figure 6-6.
7. Place the Circular Level on top of the Pedestal and check the position of the bubble.
> If the bubble is not centered, the Scale must be leveled.
> If the bubble is centered, the Scale is level. Reinstall the Circular Level in the well on top
of the DCM. Next, determine if the DCM needs leveling and height adjustment.
Item Description
1 Scale Pedestal
2 Dispensing Calibration Module
3 Scale Leveling Screws
4 Balance Mechanism
5 Circular Level
Figure 6-3 Leveling the Scale
4
1
5
3
2