3-26 Component Description
3.7.3.2 Minor Probe Height Adjustment
To make minor adjustments to the probe height:
1. Verify the following:
The probe is in the down position
The dispensing head is at the top of its Z-axis travel and at the front of the
dispensing chamber
The dispensing needle has been installed
2. Press
OFF (0) on the operator’s console and open the hatch.
3. Using a 0.050-inch hex key, loosen the micrometer locking screw. See Figure 3-14.
4. Turn the micrometer to raise or lower the probe such that the bend in the probe is visually
even with the tip of the dispensing needle as shown in Figure 3-15.
NOTE Aligning the probe bend with the needle tip automatically provides the recommended
1.78 to 2.54 mm (0.07 to 0.10 in.) height necessary for accurate height sensing.
5. After the probe height has been adjusted, tighten the micrometer locking screw and remove
the hex key.
6. If no further adjustments are necessary, close the hatch and press
ON (l) on the operator’s
console.
7. Perform a Valve Offsets routine as specified in the applicable Fluidmove User Guide (for
Windows NT or Windows XP) or Online Help.