IntellIsys AdAptIve pressure Controllers
APC-OP-LIT 1/12
Status LEDs
Valve
Open/Close
Switch
Momentary
Power
On/Off
Switch
3.2 - A/C Powered Buried Box Controllers Interface
Refer to the following diagrams and tables when connecting to an A/C powered APC Buried Box Controller including models APC-100-A, APC-200-A, APC-300-A, APC-
700-A and APC-800-A.
TABLE 3.2.6 -
(J6) “APC-X00-A”ANALOG / TTL
CONNECTOR, DB-37 RECEPTACLE
PIN FUNCTION
1 N/C
2 N/C
3 N/C
4 Digital GND / RS232 Common
5 \CDG2 Select In\
6 \Position Control Select In\
7 N/C
8 \Close Valve In\
9 \Control Active Out\
10 \Analog Div 10 Select In\
11 \Analog SP Control Select In\
12 Digital GND / RS232 Common
13 TTL Out Common
14 N/C
15 N/C
16 N/C
17 N/C
18 Analog Valve Out (0-5 VDC)
19 \Valve Open Out\
20 N/C
21 N/C
22 N/C
23 \Valve Closed Out\
24 Fault Out
25 \Zero Gauge Select In\
26 See NOTE below
27 \Open Valve In\
28 PLO #2 Status
29 PLO #1 Status
30 N/C
31 N/C
32 N/C
33 Analog SP+ In
34 Analog SP- In
35 Analog GND
36 Analog CDG Out
37 Analog Valve Out
TABLE 3.2.3 - (J5)
“APC-X00-A” VALVE
CONNECTOR,
DB-15 RECEPTACLE
PIN
SIGNAL
ASSIGNMENT
1 Drive A+
2 Drive A-
3 Drive B-
4 Drive B+
5 Sense B-
6 Sense B+
7 Sense A-
8 Sense A+
9 Reserved
10 Reserved
11 Reserved
12 Reserved
13 Reserved
14 Reserved
15 Reserved
TABLES 3.2.5 - (SW1) “APC-X00-A” DIP SWITCH
SETTINGS FOR SERIAL COMMUNICATIONS
BAUD RATE SWITCH 1 SWITCH 2
19200 OFF (up) OFF (up)
9600 ON (down) OFF (up)
4800 OFF (up) ON (down)
1200 ON (down) ON (down)
PARITY SWITCH 3 SWITCH 4
None OFF (up) Either ON or OFF
Odd ON (down) OFF (up)
Even ON (down) ON (down)
STOP BITS* SWITCH 5
1 OFF (up)
2 ON (down)
* If parity is enabled (either odd or even) then only 1 stop bit is
used and SW1 position 5 will be ignored
OTHER
FUNCTIONS
SWITCH
6
SWITCH
7
SWITCH
8
ON (down)
Enable
RS-485
Analog S.P.
0-5 VDC
Reserved
OFF (up) RS-232
Analog S.P.
0-10 VDC
Reserved
TABLE 3.2.1 - (J3 & J4)
“APC-X00-A” GAUGE
CONNECTOR DB-9 RECEPTACLE
(IDENTICAL PINOUTS)
PIN
SIGNAL
ASSIGNMENT
1 Signal input from gauge
2 N/C
3 N/C
4 +15VDC output to gauge
5 -15VDC output to gauge
6 N/C
7 N/C
8
Signal common from
gauge
9
Power common
output to gauge
TABLE 3.2.4 - (J2) “APC-X00-ARS-
232 SERIAL PORT, DB-9 RECEPTACLE
PIN
SIGNAL
ASSIGNMENT
1 RS485 (B)
2
RS-232 TX (data from
APC to host)
3
RS-232 RX (data from
host to APC)
4 N/C
5 Signal Common
6 N/C
7 N/C
8 N/C
9 RS485 (A)
TABLE 3.2.2 - (J1) “APC-X00-A” A/C
P
O
WER
C
O
NNECTOR
PIN FUNCTION
1 Neutral
2 Ground
3 Line in 100 to 240 VAC
1 3
2
VALVEVALVE
POWER
TTL/ANALOG I/O
TTL/ANALOG I/O
GAUGE 1GAUGE 1
SW1SW1
!!
100-240V~
50-60 Hz
SERIALSERIAL
GAUGE 2GAUGE 2
6.50
2.75
DEPTH: 10.60
NOTE: Function \names\ with a
backward slash indicate active low
input or output signals
NOTE: For 100-, 200- &
300- series controllers,
pin 26 is defined as
\Hold Valve Select In\
NOTE: For 700- &
800- series controllers,
pin 26 is defined as
\Valve Initialization
Enable In\