Push the <ZERO> button again:
Fine adjustment
within 0...3 s:
the zero adjustment value changes
by one unit (push <ZERO> button in
intervals of 1 s)
Change of direction
within 3...5 s:
the zero adjustment changes its
direction (the blinking frequency of
the <STATUS> LED changes
briefly)
If the <ZERO> button is released for more than 5 s, the
gauge returns to the measurement mode.
The <STATUS> LED blinks green if the signal output is negative
(< -20 mV).
4.3 Switching Functions
The two switching functions can be set to any pressure within
the measurement range of the gauge ( 14).
The current setpoint setting
is output at the D-Sub connector instead of the measurement
signal ( 20) and can be measured with a voltmeter after
the <SP> button is pressed, or
can be read/written via the RS232C interface.
If the pressure is lower than the setpoint, the corresponding LED
(<SP1> or <SP2>) is lit solid and the corresponding relay
( 20) is energized.