Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration ≤8 s)
is completed.
Press the button briefly
max. ø1.1 mm
After zero adjustment, the gauge automatically returns
to the measurement mode.
The <STATUS> LED blinks green if
the signal output is negative (< -20 mV) when the final pres-
sure has been attained
the zero adjustment has failed.
4.2.2 <ZERO> Adjustment with Ramp Function
The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.
It also permits to adjust an offset of the characteristic curve in
order to
compensate for the offset of the measuring system or
obtain a slightly positive zero for a 010 V AD converter.