12 CDG100-OP-LIT (2010-1)
Supply voltage
at the gauge
Class 2 / LPS
+14 … +30 VDC or
±15 V (±5%)
Ripple
≤1 V
pp
Power consumption
while being heated
at operating temperature
≤15 W
≤10 W
The gauge is protected against reverse polarity of the supply
voltage and overload.
Electrical connection
15-pin D-Sub, male
Sensor cable
15-pin plus shielding
Cable length
Supply voltage 15 V
≤ 4 m (0.14 mm²/conductor)
≤ 7 m (0.25 mm²/conductor)
Supply voltage 24 V
≤20 m (0.14 mm²/conductor)
≤35 m (0.25 mm²/conductor)
Supply voltage 30 V
≤40 m (0.14 mm²/conductor)
≤70 m (0.25 mm²/conductor)
For longer cables, larger conductor cross-sections are required
(R
cable
≤
1.0
Ω
).
Grounding concept
→ "Power Connection"
Materials exposed to vacuum
ceramics (Al
2
O
3
≥99.5%),
stainless steel AISI 316L,
nickel, glass solder
Internal volume
≤
4.2 cm
3
Admissible pressure (absolute)
1000 Torr (F.S.)
1 … 100 Torr (F.S.)
0.1 Torr (F.S.)
3 bar
2 bar
1.3 bar
Burst pressure (absolute) 6 bar