Product Manual

42
Technical Information
Equipment Response Time
If the signal from the sensor is inadequate, the last measured SpO
2
and pulse rate values freeze
for 10 seconds and are then replaced with dashes.
Example - SpO
2
Exponential Averaging
SpO
2
decreases 0.75% per second (7.5% over 10 seconds)
Pulse Rate = 75 BPM
Specific to this example:
The response of the 4-beat average is 1.5 seconds.
SpO
2
Values Average Latency
Standard/Fast Averaged SpO
2
4 beat exponential 2 beats
Pulse Rate Values Response Latency
Standard/Fast Averaged Pulse Rate 4 beat exponential 2 beats
Equipment Delays Delay
Display Update Delay 1.5 seconds
60
65
70
75
80
85
90
95
100
0.0
8.0
16.0
24.0
32.0
40.0
48.0
56.0
64.0
72.0
80.0
Time in seconds
SpO2
SaO2 Ref erence 4 Beat Average