Product Manual
27
Care and Maintenance
Care and Maintenance
The device requires no calibration or maintenance other than battery replacement. The device’s
expected service life is 5 years.
Cleaning the Device
Wipe the device with a soft cloth dampened with a 10% bleach/90% water solution (household
bleach [containing less than 10% sodium hypochlorite]). Do not use undiluted bleach or any
cleaning solution other than those recommended here, as permanent damage could result. Dry
with a soft cloth, or allow to air dry.
Clean once per week or more frequently if handled by multiple users.
Cleaning the Sensor
Refer to the sensor Instructions for Use for cleaning information.
Cleaning the Wristband
The wristband is designed for single-patient use. If it needs to be cleaned, hand wash with a
mild detergent (see note) in cool water (30 °C/86 °F). Allow to air dry.
Do not machine wash or dry. The wristband will shrink if placed in a dryer.
NOTES:
• Mild detergents, such as hand and dish washing liquid detergents, dissolve dirt and grease.
To clean washable surfaces, use in a solution of warm water.
• Replace the wristband if the hook and loop fastener no longer secures the wristband to the
device or to the patient.
Storing
Store the device within the stated environmental specifications. See “Specifications” for
additional information.
Remove the batteries and disconnect the sensor if it is to be stored for more than 1 month.
CAUTION: Do not place the WristOx
2,
Model 3150, in liquid or clean it with agents
containing ammonium chloride or isopropyl alcohol.
CAUTION: Use a detergent that is safe for skin and washable surfaces. Most
detergents can be high sudsing, so use sparingly. Wipe with a damp, detergent-free
cloth to remove residue.
CAUTION: After cleaning the single-patient use wristband, it should only be applied
to the same patient; do not apply it to a different patient.
!
!
!