Product Manual
24
Using the WristOx
2,
Model 3150
Activation Switch
The activation switch is located next to the sensor port at the top of the WristOx
2,
Model 3150.
It is primarily used to:
• Activate the Bluetooth radio when the device is either on or in Standby.
• Activate the device when it is in Sensor Activation mode so the user does not need to
disconnect and reconnect the sensor.
It will also activate the device when it is in Spot Check and Programmed modes.
Activate Bluetooth Radio
When the device’s Bluetooth radio is on, a master device can connect to it. If a connection is
not made, the Bluetooth radio shuts down.
Pressing the activation switch turns the Bluetooth radio on for 3 minutes. The device will remain
on until the Bluetooth radio shuts down. For example, if in Sensor Activation mode, unplugging
the sensor will not put the device in Standby.
Activate Device
When in Sensor Activation mode, the device enters Standby mode after 10 minutes without a
signal. Pressing the activation switch allows the user to turn the device on without disconnecting
and reconnecting the sensor.
Error Codes
This device includes error codes that indicate problems with the unit. When an error occurs, the
device displays the letters “Er” and a two-digit code (table 2).
Some error codes may be corrected by the user. See “Troubleshooting” for more information.
Table 2: Error Codes
Error Code Description
01 Configuration sector error
02 Patient data pointer error
03
Main memory pointer error (Device memory is intact; however, the most recent
session may be missing from the device.)
04 Data format 13 stored packet pointer error
05
Main data format 13 pointer error (Device memory is intact; however, the most
recent stored measurement may be missing from the device.)