User Guide
ALS troubleshooting
Context
• If a phototransistor is broken, replace it with a typical phototransistor.
• The phototransistor has to be also replaced, if calibration values are lost by some other reason (e.g. after
replacing the NOR memory chip D3000).
• If the phototransistor is changed, the calibration value in the phone memory has to be changed to the default
value ‘1’.
• Make sure that you have completed Display and keypad backlight troubleshooting (Page 1–38) first before
starting ALS troubleshooting.
Here are some hints for ALS troubleshooting; the following troubleshooting diagram refers to these:
• Phoenix LED control tool also shows you luminance. The correct luminance in darkness is <20lx, and in office
environment 100-2000lx. The luminance value depends strongly on the light source and the angle of the
phone, so these values are only a rough guideline.
• LED driver control voltage measurement points can be found from LED driver troubleshooting (Page
1–41) section. When backlight brightness is set to 100%, both GENOUT-signals are low, and enable PWM is
100%.
• Phoenix has an ambient light sensor calibration tool for changing calibration values. The pull-up resistor
calibration is done first:
a Cover the light guide (upper part of the A Cover).
b Click Start Write .
c Manually change the ambient light sensor value to the default value. There is no special tool for this, but
you have to perform calibration normally and then set the Co-efficient result to 1 before writing it to the
phone memory.
Figure 16 Ambient Light Sensor Calibration window
RM-84/99
BB Troubleshooting and Manual Tuning Guide Nokia Customer Care
Issue 1 Company Confidential Page 1–39
Copyright ©2005 Nokia. All Rights Reserved.










