13 To complete the procedure, click “Finish”.
FLS-4 can be used right after the flash update package is installed.
FPS-8* flash prommer must be updated by using Phoenix!
Updating FPS-8* flash prommer SW
To update the flash prommer software:
1 Start Phoenix service software and login.
Manage connection correctly for the FPS-8* flash prommer.
2 From the “Flashing” menu, choose ”FPS-8 Maintenance”.
3 When a new FPS-8 flash update package is installed to your computer, you are
asked to update the files to your FPS-8 Prommer.