Adjust the level of signal CHF to minimum by varying D/A converter value
with cursor left and right keys or <+> and <–> keys.
When value is OK move to the next option with cursor keys.
Select option ”TX I and Q DC offset” .
Adjust the level of signal CHF to minimum by varying D/A converter value
with cursor keys or <+> and <–> keys.
When value is OK move to the next option with cursor keys.
Select option ”Tune Amplitude Difference”.
Adjust the level of signal CHF+67.7 kHz to minimum by varying D/A convert-
er value with cursor keys or <+> and <–> keys.
The amplitude difference between CHF+67.7 kHz and CHF–67 kHz should
be >35 dB (See picture).
When value is OK move to the next option with cursor keys.
Select option ”Tune Phase Difference”.
Adjust the level of signal CHF+67.7 kHz to minimum by varying D/A convert-
er value with cursor left and right keys or <+> and <–> keys.
When spectre is minimized press <Esc> key and store new values to EE-
PROM with <Y> key.
Press <ESC> key to return to upper menu (Testing and Adjustment).
Select option ”RF Controls”.
Change ”TX Data Type” value from one to zero or from zero to one depend-
ing on wich value the option had.
Press <ESC> key to return to upper menu (Testing and Adjustment).
Select option ”Adjustment”.
Select option ”TX I/Q Tuning”.
Press <Y> key to load values from EEPROM.
Minimize the spectre like explained earlier. Check the other side of the
spectre once more.
When both sides of the spectre are minimized, press <Esc> key and store
new values to EEPROM with <Y> key.