User's Manual
Withings Blood Pressure Monitor User’s Guide
15
RFStatement
‐
MedicalElectricalEquipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledand
putintoserviceaccordingtotheEMCinformationprovidedinthefollowing.
‐Interferencemayoccurinthevicinityofequipmentmarkedwith
‐PortableandmobileRFcommu nicationequipment(e.g. cellphones)canaffectMedicalElectrical
Equipment.
‐Theuseofa
ccessoriesandcablesotherthanthosespecifiedmayresultinincre asedemissions or
decreasedimmunity.
‐ThedeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylowand
arenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
‐Thedeviceissuitableforuseinallesta
blishments,includingdomesticestablishmentsandthose
directlyconnectedtothepubliclowvoltagepowersupplynetworkthatsuppliesbuildingsusedfor
domesticpurposes.
‐PortableandmobileRFcommu nicationsequipmentshouldbeusednoclosertoanypartofthe
device,includingcables,thantherecommendedseparationdistan
cecalculatedfromtheequation
applicabletothefre quencyofthetransmitter.
‐ThedeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdistu rbances
arecontrolled.Anyotheraccessories,transducersandcablesmayresultinincreasedemissionsor
decreasedimmunityandEMCperformance.
‐Thedevicesh
ouldnotbeusedadjacenttoorstackedwithotherequipmentandthatifadjacentor
stackeduseisnecessary,whichshouldbeobservedtoverifynormaloperationin theconfiguration
inwhichitwillbeused.
MedicalElectricalEquipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledand
putintoservicea
ccordingtotheEMCinformationprovidedinthefollowing.PortableandmobileRF
communicationequipment(e.g.cellphones)canaffectMedicalElectricalEquipment.Theuseof
accessoriesandcablesotherthanthosespecifiedmayresultinincreasedemissionsordecreased
immunityoftheunit.
Guidanceandmanufacturer’sdeclaration‐electromagneticemissions
TheWirelessBloodPressureMonitorisintendedforuseintheelectromagnetic environment
specifiedbelow.ThecustomerortheuseroftheWirelessBloodPressureMonitorshouldassure
thatitisusedinsuchanenvironment.
Emissionstest Compliance Electromagneticenvironment‐guidance
CEemissions
CISPR11
Group1
TheBP‐801WirelessBloodPressureMonitorusesRF
energyonlyforitsinternalfunction.
Therefore,itsRFemissionsareverylowandarenot
likelytocauseanyinterferencein
nearbyelectronicequipment.
REemissions
CISPR11
ClassB
Harmonic
emissions
IEC61000‐3‐2
Notapplicable
TheBP‐801WirelessBloodPressureMonitorissuitable
foruseinallestablishments,includingdomestic
establishmentsandthosedirectlycon nectedtothe
publiclow‐voltagepower
supplynetworkthatsuppliesbuildingsusedfor
domesticpurposes.
Voltage
fluctuations/
Flickeremissions
IEC61000‐3‐3
Notapplicable