User's Manual

3
. Replace the AAA alkaline
cells
“Low bat
t
ery
icon
Battery is low
Replace batteries.
“Error”
The temperature is outside
ranges.
Body mode : lower than 35°C
(95°F) or higher than 43.2°C
(109.76°F)
Surface mode : lower than
22°C (71.6°F) or higher than
40°C (104°F)
1.
Make a new
temperature
measurement,
referring to the user
manual.
“Error”
System error
Contact customer service
The temperature seems
rather
high
The unit has been stored in a
cold room
Leave the unit in the room
where is the patient for 30
minutes prior to the next
measurement
The temperature seems rather
low
The skin has sweat on it or
there is hair on the skin
Ensure that there is no
haire in
front of the sensor. Clean the
skin with a dry piece of cloth
and wait for 5 minutes prior to
taking a measurement
Measurement was
taken
in
surface mode
Switch in body mode (see
manual)
The patient has been staying in
a cold room
Wait until
the patient is
warmer before proceeding to
measurement
Measurement was not
taken
on the temple
Please refer to the user manual
to place the sensor on the
correct location
Bluetooth does not seem to
work
Smartphone is out of range
Please get your smartph
one
closer to your device.
Smartphone’s Bluetooth is OFF
Please switch your smartphone
Bluetooth ON
Wi
-
Fi does not seem to work
Device is out of range from the
Wi-Fi source
Please get your smartphone
and device closer to your
device.
Wi
-
Fi is OFF
Please switch your smartphone
Wi-Fi ON as well as your router
Wi-Fi on