Test Set Calibration – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP 3-103
PRELIMINARY
Detecting Test Equipment
when using Agilent E7495A
Verify that no other equipment is connected to the LMF.
IMPORTANT
*
Agilent E7495A equipment must be connected to the LAN to detect it.
Then perform the procedures described in Appendix F. Specifically, refer
to Table F-1 on Page F-1, Table F-2, and Table F-3 on Page F-3.
Calibrating Test Equipment
The LMF Calibrate Test Equipment Procedure zeros the Power
Measurement Level of the Test Equipment Item that is to be used for TX
Calibration and Audit. If both a Power Meter and an Analyzer are
connected (for example, an HP 437 and an HP 8921A/600), only the
Power Meter is Zeroed.
The Agilent E4406A Transmitter Tester does not support
Power Measurement Level zeroing.
– Refer to the Equipment Calibration section of
Appendix F for Agilent E4406A Calibration.
NOTE
Prerequisites
S LMF Computer Serial Port and Test Equipment are connected to the
GPIB Box.
S Test Equipment is turned on and has warmed–up for at least 60
minutes.
S Test Equipment has been selected in the LMF (Table 3-29 or
Table 3-30)
Perform the procedure in Table 3-31 to calibrate the Test Equipment.
Table 3-31: LMF Test Equipment Calibration Procedure
n Step Action
1 From the Util Menu, select Calibrate Test Equipment from the Pull–down Menu.
–A Directions Window is displayed.
2 Follow the directions provided.
3 Click on Continue to close the Directions Window and start the Calibration Process.
–A Status Report Window is displayed.
4 Click on OK to close the Status Report Window.