User's Manual

Table Of Contents
Power Delta Calibration – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP H-5
PRELIMINARY
Table H-1: Agilent E4406A Power Delta Calibration Procedure
n ActionStep
9i Using the Softkeys, select External Atten and then select Mobile.
9j Using the Numeric Keypad, set Mobile to 0 (zero) and, using the Softkeys, select dB.
9k Using the Softkeys, select Base.
9l Using the Numeric Keypad, set Base to 0 (zero) and, using the Softkeys, select dB.
9m Press MEASURE and, using the Softkeys, select Channel Power.
10 On the E4432B Signal Generator, press RF On/Off to toggle the RF Output to RF ON.
Note that the RF On/Off status in the Screen Display changes.
11 Read the measured Channel Power from the E4406A Screen Display and record it as Result B.
B____________________.
12 On the E4432B, press RF On/Off to toggle the RF Output to RF OFF.
Note that the RF On/Off status in the Screen Display changes.
13 Calculate the Power Calibration Delta Value.
The Delta Value is the Power Meter Measurement minus the Agilent measurement.
Delta = A – B
Example: Delta = –0.70 dBm – (–1.25 dBm) = 0.55 dBm
Example: Delta = 0.26 dBm – 0.55 dBm = –0.29 dBm
NOTE
These examples are included to show the mathematics and do not represent actual readings.
NOTE
Add this Delta Value to the TX Cable Loss Value during In–Service Calibration.
Refer to Step 4 in Table H-6.
Power
Sensor
Agilent E4432B and E4406A
Short RF Cable
HP437B
SENSOR
RF OUTPUT
Figure H-1: Delta Calibration Set–up – Agilent E4432B to HP437
H