Calibrating the Gigatronics
8541C Power Meter
Precise transmit Output Power Calibration measurements are made using
a Bolometer–type Broadband Power Meter with a sensitive Power
Sensor. Perform the procedure in Table F-22 to enter information unique
to the Power Sensor.
Table F-22: Gigatronics 8541C Power Meter Calibration Procedure
n Step Action
! CAUTION
Do not connect/disconnect the Power Meter Sensor Cable while AC Power is applied to the meter.
Disconnection could result in destruction of the Sensing Element or miscalibration.
1 Make sure the Power Meter POWER Pushbutton is OFF.
3 Connect the Power Sensor Cable to the SENSOR Input.
4
Set the POWER Pushbutton to ON.
NOTE
Allow the Power Meter and sensor to warm–up and stabilize for a minimum of 60 minutes before
performing the Calibration Procedure.
5 Connect the Power Sensor to the CALIBRATOR Output Connector.
6 Press ZERO.
Wait for the process to complete.
Sensor Factory Calibration Data is read to Power Meter during this process.
7 When the Zeroing Process is complete, disconnect the Power Sensor from the CALIBRATOR
Output.
CONNECT POWER SENSOR
WITH POWER METER
TURNED OFF
CONNECT POWER SENSOR TO
CALIBRATOR POWER REFERENCE
WHEN CALIBRATING/ZEROING UNIT
FRONT View REAR View
GPIB CONNECTIONAC POWER
FW00564
Figure F-22: Gigatronics 8541C Power Meter Detail