Equipment Calibration – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-33
PRELIMINARY
Table F-21: HP 437 Power Meter Calibration Procedure
n ActionStep
11 Perform the following actions to set the CAL FAC %:
11a Press [SHIFT] then [FREQ] for CAL FAC.
11b On the Sensors Decal, locate an approximate Calibration Percentage Factor (CF%) at 2 GHz.
11c Enter the Sensors Calibration % (CF%) using the Arrow Keys and press [ENTER].
When complete, the Power Meter will typically display 0.05 dBm; any reading between 0.00
and 0.10 is normal.
12 To turn off the PWR REF, perform the following actions.
12a Press [SHIFT] then [].
12b Disconnect the Power Sensor from the POWER REF Output.
F