Verifying and Setting GPIB Addresses – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-17
PRELIMINARY
Gigatronics 8541C Power
Meter GPIB Address
Refer to Figure F-11 and follow the steps in Table F-12 to verify and, if
necessary, change the Gigatronics 8541C Power Meter GPIB Address.
MENU ENTER ARROW
KEYS
1
REF FW00564
Figure F-11: Gigatronics 8541C Power Meter Detail
This procedure assumes that the Test Equipment is set–up
and ready for testing.
NOTE
Table F-12: Verify and/or Change Gigatronics 8541C Power Meter GPIB Address
n Step Action
1
! CAUTION
Do not connect/disconnect the Power Meter Sensor Cable with AC Power applied to the meter.
Disconnection could result in destruction of the Sensing Element or miscalibration.
Press MENU.
2 Use the b Arrow Key to select CONFIG MENU and press ENTER.
3 Use the b Arrow Key to select GPIB and press ENTER.
The current Mode and GPIB Address are displayed.
4 If the Mode is not set to 8541C, perform the following to change it:
Use the a Arrow Keys as required to select MODE.
Use the by Arrow Keys as required to set MODE to 8541C.
5 If the GPIB Address is not set to 13, perform the following to change it:
Use the Arrow Key to select ADDRESS.
Use the by Arrow Keys as required to set the GPIB Address to 13.
6 Press ENTER to return to normal operation.