Verifying and Setting GPIB Addresses – continued
F-8 1X SC 4812T Lite BTS Optimization/ATP FEB 2005
PRELIMINARY
Advantest R3267 Spectrum
Analyzer GPIB Address
Refer to Figure F-5 and perform the procedure in Table F-6 to verify
and, if necessary, change the Advantest R3267 Spectrum Analyzer GPIB
Address.
o
n
REMOTE LED
LCL Key
CONFIG Key
Softkey Label Display Area
Softkey Buttons
Keypad BS Key ENTR Key
Figure F-5: Setting Advantest R3267 GPIB Address
Table F-6: Verify that and Change Advantest R3267 GPIB Address
n Step Action
1 If the REMOTE LED is lighted, press the LCL Key.
The LED extinguishes.
2 Press the CONFIG Key.
CONFIG Softkey Labels will appear in the Softkey Label display area of the instrument
display.
The current GPIB Address will be displayed below the GPIB Address Softkey Label.
3 If the current GPIB Address is not set to 18, perform the following to change it:
3a Press the GPIB Address Softkey.
–A GPIB Address Entry window will open in the instrument display showing the current
GPIB Address.
table continued on next page