Verifying and Setting GPIB Addresses
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-5
PRELIMINARY
Agilent E4406A Transmitter
Tester GPIB Address
Refer to Figure F-3 and perform the procedure in Table F-4 to verify
and, if necessary, change the Agilent E4406A GPIB Address.
System Key
Bk Sp
Key
Enter
Key
Data Entry Keypad
Softkey Buttons
Softkey Label Display AreaActive Function Area
ti-CDMA-WP-00085-v01-ildoc-ftw
Figure F-3: Setting Agilent E4406A GPIB Address
Table F-4: Verify and Change Agilent E4406A GPIB Address Procedure
n Step Action
1 In the SYSTEM section of the Instrument Front Panel, press the System Key.
The Softkey Labels displayed on the right side of the instrument screen changes.
2 Press the Config I/O Softkey Button to the right of the instrument screen.
The Softkey Labels changes.
The current instrument GPIB Address will be displayed below the GPIB Address Softkey Label.
3 If the current GPIB Address is not set to 18, perform the following actions to change it.
3a Press the GPIB Address Softkey Button.
In the on–screen Active Function Area, GPIB Address will be displayed followed by the
current GPIB Address.
table continued on next page