User Manual
GPIB
68P09255A57-2
Aug 2002
SC4812ET Optimization/ATP Manual Software Release R16.1.x.x
PRELIMINARY
I-12
Figure I-9: Setting Agilent E4406A GPIB Address
System
Key
Bk Sp
Key
Enter
Key
Data Entry
Keypad
Softkey
Buttons
Softkey Label
Display Area
Active Function
Area
Agilent E4432B Signal Generator GPIB Address
Follow the procedure in Table I-9 and refer to Figure I-10 to verify and,
if necessary, change the Agilent E4432B GPIB address.
Table I-9: Verify and Change Agilent E4432B GPIB Address
Step Action
1 In the MENUS section of the instrument front panel, press the Utility key.
- The softkey labels displayed on the right side of the instrument screen will change.
2 Press the GPIB/RS232 softkey button to the right of the instrument screen.
- The softkey labels will change.
- The current instrument GPIB address will be displayed below the GPIB Address softkey label.
3 If the current GPIB address is not set to 1, perform the following to change it:
3a - Press the GPIB Address softkey button.
-- The GPIB Address label and current GPIB address will change to boldface.
-- In the on-screen Active Entry Area, Address: will be displayed followed by the current
GPIB address.
. . . continued on next page