Equipment Calibration – continued
B-30 1X SC480 BTS Hardware Installation, Optimization/ATP, and FRU Jun 2004
DRAFT
Table B-21: HP 437 Power Meter Calibration Procedure
Step Action
5a Observe the instrument display and determine if the triangular indicator over PWR REF is
displayed.
5b If the triangular indicator is displayed, press [SHIFT] then [] to turn it off.
6 Press [ZERO].
Display will show “Zeroing ******.”
Wait for process to complete.
7 Connect the power sensor to the POWER REF output.
8 Turn on the PWR REF by performing the following:
8a Press [SHIFT] then [].
8b Verify that the triangular indicator (below) appears in the display above PWR REF.
9 Perform the following to set the REF CF%:
9a Press ([SHIFT] then [ZERO]) for CAL.
9b Enter the sensors REF CF% from the sensors decal using the arrow keys and press [ENTER].
(The power meter will display ”CAL *****” for a few seconds.)
NOTE
If the REF CAL FACTOR (REF CF) is not shown on the power sensor, assume it to be 100%.
10 Perform the following to set the CAL FAC %:
10a Press [SHIFT] then [FREQ] for CAL FAC.
10b On the sensor’s decal, locate an approximate calibration percentage factor (CF%) at 2 GHz.
10c Enter the sensors calibration % (CF%) using the arrow keys and press [ENTER].
–– When complete, the power meter will typically display 0.05 dBm. (Any reading between
0.00 and 0.10 is normal.)
11 To turn off the PWR REF, perform the following:
11a Press [SHIFT] then [].
11b Disconnect the power sensor from the POWER REF output.