F-22 1X SC4812T–MC BTS Optimization/ATP 5/21/04
DRAFT
Agilent E4406A Transmitter
Tester Self-alignment
(Calibration)
Refer to Figure F-14 and follow the procedure in Table F-16 to perform
the Agilent E4406A self–alignment (calibration).
Figure F-14: Agilent E4406A
System
Key
Softkey
Buttons
Softkey Label
Display Area
ti-CDMA-WP-00080-v01-ildoc-ftw
Table F-16: Perform Agilent E4406A Self–alignment (Calibration)
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
– The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Alignments softkey button to the right of the instrument screen.
– The softkey labels will change.
3 Press the Align All Now softkey button.
– All other instrument functions will be suspended during the alignment.
– The display will change to show progress and results of the alignments performed.
– The alignment will take less than one minute.
HP 437 Power Meter
(Calibration)
Precise transmit output power calibration measurements are made using
a bolometer–type broadband power meter with a sensitive power sensor.
Follow the steps in Table F-17 to enter information unique to the power
sensor before calibrating the test setup. Refer to Figure F-15 as required.