User Manual
Verifying and Setting GPIB Addresses – continued
F-14 1X SC4812T–MC BTS Optimization/ATP 5/21/04
DRAFT
Table F-11: Verify and/or Change HP437 Power Meter GPIB Address
Step Action
1 Press Shift and PRESET.
2 Use the y arrow key to navigate to HP–IB ADRS and press ENTER.
The HP–IB address is displayed.
NOTE
HP–IB is the same as GPIB.
3 If the current GPIB address is not set to 13, perform the following to change it:
– Use the y b arrow keys to change the HP–IB ADRS to 13.
– Press ENTER to set the address.
4 Press Shift and ENTER to return to a standard configuration.
Gigatronics 8541C Power
Meter GPIB Address
Follow the steps in Table F-12 and refer to Figure F-11 to verify and, if
necessary, change the Gigatronics 8541C power meter GPIB address.
Figure F-11: Gigatronics 8541C Power Meter Detail
MENU ENTER ARROW
KEYS
REF FW00564
1
This procedure assumes that the test equipment is set up
and ready for testing.
NOTE
Table F-12: Verify and/or Change Gigatronics 8541C Power Meter GPIB Address
Step Action
! CAUTION
Do not connect/disconnect the power meter sensor cable with AC power applied to the meter.
Disconnection could result in destruction of the sensing element or miscalibration.
1 Press MENU.
F