Test Equipment Calibration continued
DRAFT
SC300 1X BTS Hardware Installation, ATP, and FRU Procedures
JAN 2002
7-54
Procedure to Calibrate Agilent
E4406A
Refer to Figure 7-27 and follow the procedure in Table 7-32 to perform
the Agilent E4406A selfalignment (calibration).
Table 7-32: Perform Agilent E4406A Selfalignment (Calibration)
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Alignments softkey button to the right of the instrument screen.
The softkey labels will change.
3 Press the Align All Now softkey button.
All other instrument functions will be suspended during the alignment.
The display will change to show progress and results of the alignments performed.
The alignment will take less than one minute.
Figure 7-27: Performing Agilent E4406A Selfalignment (Calibration)
System
Key
Softkey
Buttons
Softkey Label
Display Area
7