User's Manual

GPIB Addresses continued
JAN 2002
7-47
SC300 1X BTS Hardware Installation, ATP, and FRU Procedures
DRAFT
Verify and Set Advantest R3562
Signal Generator GPIB
Address
Set the GPIP ADDRESS switch on the rear of the Advantest R3562
signal generator to address 1 as shown in Figure 7-24.
Figure 7-24: Advantest R3562 GPIB Address Switch Setting
12345678
54321
GPIP ADDRESS
1
0
GPIB Address set to 1
Verify and Set Agilent E4406A
Transmitter Tester GPIB
Address
Follow the procedure in Table 7-27 and refer to Figure 7-25 to verify
and, if necessary, change the Agilent E4406A GPIB address.
Table 7-27: Verify and Change Agilent E4406A GPIB Address
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Config I/O softkey button to the right of the instrument screen.
The softkey labels will change.
The current instrument GPIB address will display below the GPIB Address softkey label.
3 If the current GPIB address is not set to 18, perform the following to change it:
3a Press the GPIB Address softkey button.
In the onscreen Active Function Area, GPIB Address will be displayed followed by the current
GPIB address.
. . . continued on next page
7