Specifications

1X UBS Macro B T S Optimization/A TP T est Equipment Calibr ation Gigatronics 8542 P ower Meter
Procedure D -10 Calibr ate Gigatronics 8542 P ower Meter (Continued)
4
Connect the power sensor to the CALIBRA TOR output connector .
5
Press ZERO , and wait for the process to complete.
Result: Sensor factory calibration data is read to power meter during this process.
6
When the zeroing process is complete, disconnect the power sensor from the
CALIBRA TOR output.
68P09283A63 -5 D -23
A UG 2007 FOA