Specifications
T est Equipment Calibr ation – Gigatronics 8542 P ower Meter Appendix D: T est Equipment Prepar ation
Test Equipment Calibration – Gigatronics 8542 Power
Meter
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
■
■
Precise transmit output power calibration measurements are made using a bolometer -type
broadband power meter with a sensitive power sensor . F ollow the steps in Procedure D -10 to
enter information unique to the power sensor . Refer to Figure D -12 as necessary .
Figure D -12 Gigatronics 8541C P ower Meter Calibr ation
ti-cdma-00128.eps
CONNECT POWER SENSOR
WITH POWER METER
TURNED OFF
CONNECT POWER SENSOR TO
CALIBRATOR POWER REFERENCE
WHEN CALIBRATING/ZEROING UNIT
CALIBRATOR
ZERO
CAL
FREQ
REL
MENU
ESCAPE
dB/MWRECALL
LOCAL
ENTER
A
MODE
GPIB
POWER
I
O
Giga-tronics8542CUniversalPowerMeter
dB m
- 1 0 0
LINEVOLTAGE
SELECTION
GPIB
120VAC
FRONT View REAR View
GPIB CONNECTIONAC POWER
A
B
Procedure D -10 Calibr ate Gigatronics 8542 P ower Meter
1
Do not connect/disconnect the power meter sensor cable with AC power applied
to the meter . Disconnection could result in destruction of the sensing element
or miscalibr ation.
Make sure the power meter POWER pushbutton is OFF .
2
Connect the power sensor cable to the SENSOR input.
3
Set the POWER pushbutton to ON .
Allow the power meter and sensor to warm up and stabilize for a minimum of
60 minutes before performing the calibration procedure.
D -22 68P09283A63 -5
FOA A UG 2007