Specifications

1X UBS Macro B T S Optimization/A TP T est Equipment Calibr ation Agilent 4406 Self -alignment
Test Equipment Calibration Agilent 4406
Self -alignment
Refer to Figure D -11 when performing the following.
Figure D -11 Agilent E4406A Self -alignment
System
Key
Softkey
Buttons
Softkey Label
Display Area
ti-cdma-00126.eps
Procedure D -9 Agilent E4406A Self -alignment (Calibr ation)
1
In the SYSTEM section of the instrument front panel, press the System key .
Result: The softkey labels displayed on the right side of the instrument screen will
change.
2
Press the Alignments softkey button to the right of the instrument screen.
Result: The softkey labels will change.
3
Press the Align All Now softkey button.
Result:
All other instrument functions will be suspended during the alignment.
The display will change to show progress and results of the alignments performed.
The alignment will take less than one minute.
68P09283A63 -5 D -21