Specifications

V erif ying and Setting GPIB Addresses Agilent E4418 P ower Meter Appendix D: T est Equipment Prepar ation
Verifying and Setting GPIB Addresses Agilent E4418
Power Meter
Figure D -9 Agilent E4418B P ower Meter V erif y and Set/Change GPIB Address
ti-cdma-05962.eps
LEFT DOWN UP RIGHT
ARROW
KEYS
PRESET/
LOCAL
SYSTEM/
INPUTS
STATUS
REPORTING
LINE
SOFTKEY
LABEL
AREA
SOFTKEY
MENU
TITLE
SOFTKEY
MENU PAGE
NUMBER
Refer to Figure D -9 when performing the following:
Procedure D -8 V erif y and Set/Change Agilent E4418 P ower Meter GPIB Address
1
If RMT (remote operation) is displayed on the left end of the instrument screen status
reporting line, press the Preset/Local front panel button.
Result: The status line entry will change to LCL (local or front panel operation).
2
On the instrument front panel, press the System/Inputs key .
Result: The softkey labels displayed on the right side of the instrument screen will
change.
3
Press the Remote Interface > Congure Interface > GPIB softkeys to the right of the
instrument screen.
Result:
The softkey labels will change.
The current instrument GPIB address will be displayed below the GP -IB Addr
softkey label.
D -18 68P09283A63 -5
FOA A UG 2007