Refer to Figure D -8 when performing the following:
This procedure assumes that the test equipment is set up and ready for testing.
Procedure D -7 V erif y and Set/Change Gigatronics 8541C P ower Meter GPIB Address
1
Do not connect/disconnect the power meter sensor cable with AC power applied
to the meter . Disconnection could result in destruction of the sensing element
or miscalibr ation.
Press MENU .
2
Use the [Symbol_arrowblackdown] arrow key to select CONFIG MENU and press ENTER .
3
Use the [Symbol_arrowblackdown] arrow key to select GPIB and press ENTER .
Result: The current Mode and GPIB Address are displayed.