Specifications
V erif ying and Setting GPIB Addresses – Agilent E4432B Signal Gener ator Appendix D: T est Equipment Prepar ation
Procedure D -4 V erif y and Change Agilent E4432B Signal Gener ator GPIB Address
(Continued)
The GPIB Address label and current GPIB address will change to boldface.
In the on-screen Active Entry Area, Address: will be displayed followed by the
current GPIB address.
2. On the front panel Numeric keypad, enter the signal generator GPIB address of 1 .
Result:
The GPIB Address label will change to Enter .
Digits entered with the keypad will replace the current GPIB address in the
Active Entry display .
T o correct an entry , press the backspace key at the lower right of the
keypad to delete one character at a time.
3. Press the Enter softkey button to set the new GPIB address.
Result:
The new GPIB address will be displayed under the GPIB Address softkey label.
D -10 68P09283A63 -5
FOA A UG 2007