Refer to Figure D -4 when performing the following:
Procedure D -4 V erif y and Change Agilent E4432B Signal Gener ator GPIB Address
1
In the MENUS section of the instrument front panel, press the Utility key .
Result:
• The softkey labels displayed on the right side of the instrument screen will change.
2
Press the GPIB/RS232 softkey button to the right of the instrument screen.
Result:
• The softkey labels will change.
• The current instrument GPIB address will be displayed below the GPIB Address
softkey label.
3
If the current GPIB address is not set to 1 , perform the following to change it:
1. Press the GPIB Address softkey button.
Result:
68P09283A63 -5 D -9