Specifications
1X UBS Macro B T S Optimization/A TP V erif ying and Setting GPIB Addresses – Agilent E4406A T r ansmitter T ester
Verifying and Setting GPIB Addresses – Agilent E4406A
Transmitter Tester
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
■
■
Figure D -3 Setting Agilent E4406A GPIB Address
ti-cdma-00118.eps
System Key
Bk Sp Key
Enter Key
Data Entry KeypadSoftkey Buttons
Softkey Label Display AreaActive Function Area
ti-CDMA-WP-00085-v01-ildoc-ftw
Refer to Figure D -3 when performing the following:
Procedure D -3 V erif y and Set/Change Agilent E4406A T r ansmitter T ester GPIB
Address
1
In the SYSTEM section of the instrument front panel, press the System key .
Result: The softkey labels displayed on the right side of the instrument screen will
change.
2
Press the Config I/O softkey button to the right of the instrument screen.
Result:
• The softkey labels will change.
• The current instrument GPIB address will be displayed below the GPIB Address
softkey label.
3
If the current GPIB address is not set to 18 , perform the following to change it:
1. Press the GPIB Address softkey button.
Result: In the on-screen Active F unction Area, GPIB Address will be displayed
followed by the current GPIB address.
68P09283A63 -5 D -7
A UG 2007 FOA