User's Manual

Test Equipment Set-up68P09258A31–A
Oct 2003
1X SCt 4812T BTS Optimization/ATP
F-27
Agilent E4406A Transmitter Tester Self-alignment (Calibration)
The Agilent E4406A requires pre–calibration actions or calibration
verification that are not supported by the LMF. Follow the procedure in
Table F-17 and refer to Figure F-15 to perform the Agilent E4406A
self–alignment (calibration).
Table F-17: Agilent E4406A Self–alignment (Calibration)
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Alignments softkey button to the right of the instrument screen.
The softkey labels will change.
3 Press the Align All Now softkey button.
All other instrument functions will be suspended during the alignment.
The display will change to show progress and results of the alignments performed.
The alignment will take less than one minute.
Figure F-15: Agilent E4406A Self–alignment
System
Key
Softkey
Buttons
Softkey Label
Display Area
ti-CDMA-WP-00080-v01-ildoc-ftw
F