Test Equipment Set-up68P09258A31–A
Oct 2003
1X SCt 4812T BTS Optimization/ATP
F-27
Agilent E4406A Transmitter Tester Self-alignment (Calibration)
The Agilent E4406A requires pre–calibration actions or calibration
verification that are not supported by the LMF. Follow the procedure in
Table F-17 and refer to Figure F-15 to perform the Agilent E4406A
self–alignment (calibration).
Table F-17: Agilent E4406A Self–alignment (Calibration)
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
– The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Alignments softkey button to the right of the instrument screen.
– The softkey labels will change.
3 Press the Align All Now softkey button.
– All other instrument functions will be suspended during the alignment.
– The display will change to show progress and results of the alignments performed.
– The alignment will take less than one minute.
Figure F-15: Agilent E4406A Self–alignment
System
Key
Softkey
Buttons
Softkey Label
Display Area