User's Manual

Verifying and Setting GPIB Address68P09258A31–A
Oct 2003
1X SCt 4812T BTS Optimization/ATP
F-15
Gigatronics 8541C Power Meter GPIB Address
Refer to Figure F-11 and follow the steps in Table F-12 to verify and, if
necessary, change the Gigatronics 8541C power meter GPIB address.
Figure F-11: Gigatronics 8541C Power Meter Detail
MENU ENTER ARROW
KEYS
1
REF FW00564
NOTE
This procedure assumes that the test equipment is set up and
ready for testing.
Table F-12: Verify and/or Change Gigatronics 8541C Power Meter GPIB Address
Step Action
1
! CAUTION
Do not connect/disconnect the power meter sensor cable with AC power applied to the meter.
Disconnection could result in destruction of the sensing element or miscalibration.
Press MENU.
2 Use the b arrow key to select CONFIG MENU and press ENTER.
3 Use the b arrow key to select GPIB and press ENTER.
The current Mode and GPIB Address are displayed.
4 If the Mode is not set to 8541C, perform the following to change it:
Use the a arrow keys as required to select MODE.
Use the by arrow keys as required to set MODE to 8541C.
5 If the GPIB address is not set to 13, perform the following to change it:
Use the arrow key to select ADDRESS.
Use the by arrow keys as required to set the GPIB address to 13.
6 Press ENTER to return to normal operation.