User's Manual

Test Equipment Preparation
68P09258A31–A
Oct 2003
1X SCt 4812T BTS Optimization/ATP
F-4
Detecting Test Equipment
Check that no other equipment is connected to the LMF. Agilent
equipment must be connected to the LAN to detect it. Then perform the
procedures described in Table F-2.
Table F-2: Detecting Agilent E7495A Test Equipment
n Step Action
1 Click the Tools Menu.
2 Choose Options.
3 Check Agilent E7495A option in non–GPIB Test Equipment and enter its IP number.
4 Click Apply and wait a moment.
5 Click Dismiss.
Power Sensor Calibration
Table F-3 describes the E7495A Power Sensor Calibration.
Table F-3: E7495A Power Sensor Calibration
n Step Action
1 Display the power meter screen.
2 Zero the power meter. Make sure you are connected as shown in Figure F-1.
Press the Zero softkey.
Press the Continue softkey.
3 Calibrate the power meter:
Press Ref CF.
Enter the reference cal factor, reading it off the label on the power sensor head.
Press Calibrate.
Connect the power sensor (see Figure F-2).
Press Continue.
Press Cal Factor.
Enter the cal factor from the label on the power sensor head. Select a cal factor that’s within
the operating frequency of the base station.
F