User's Manual Part 4
Test Equipment Connection, Testing, and Control – continued
MAY 2004 1X SC480 BTS Hardware Installation, Optimization/ATP, and FRU B-27
PRELIMINARY
Advantest R3267/R3562 Test
Equipment Interconnection
To provide proper operation during testing when both units are required,
the R3257 spectrum analyzer must be interconnected with the R3562
signal generator as follows:
10 MHz reference signal – Connect a BNC (M)–BNC (M) cable
between the R3562 SYNTHE REF IN connector and the R3267 10
MHz OUT connector as shown in Figure B-19.
Serial I/O – Using the Advantest cable provided, connect the R3267
SERIAL I/O connector to the R3562 SERIAL I/O connector as shown
in Figure B-19.
TDME0010–1
R3562
SYNTHE REF IN
TO GPIB
BOX
R3562
SERIAL I/O
TO GPIB BOX
R3267
10 MHZ OUT
R3267
SERIAL I/O
Figure B-19: Advantest 10 MHz Reference and Serial I/O Connections
B