User's Manual Part 4

Verifying and Setting GPIB Addresses – continued
MAY 2004 1X SC480 BTS Hardware Installation, Optimization/ATP, and FRU B-7
PRELIMINARY
Table B-5: Verify and Change Agilent E4432B GPIB Address
Step Action
1 In the MENUS section of the instrument front panel, press the Utility key.
The softkey labels displayed on the right side of the instrument screen will change.
2 Press the GPIB/RS232 softkey button to the right of the instrument screen.
The softkey labels will change.
The current instrument GPIB address will be displayed below the GPIB Address softkey label.
3 If the current GPIB address is not set to 1, perform the following to change it:
3a Press the GPIB Address softkey button.
The GPIB Address label and current GPIB address will change to boldface.
In the on–screen Active Entry Area, Address: will be displayed followed by the current GPIB
address.
3b
On the front panel Numeric keypad, enter the signal generator GPIB address of 1.
The GPIB Address label will change to Enter.
Characters typed on the keypad will replace the current GPIB address in the Active Entry display.
NOTE
To correct an entry, press the backspace key at the lower right of the keypad to delete one character at
a time.
3c Press the Enter softkey button to set the new GPIB address.
The new GPIB address will be displayed under the GPIB Address softkey label.
Advantest R3267 Spectrum
Analyzer GPIB Address
Refer to Figure B-5 and perform the procedure in Table B-6 to verify
and, if necessary, change the Advantest R3267 spectrum analyzer GPIB
address.
B