User's Manual Part 2

Test Set Calibration – continued
MAY 2004 1X SC480 BTS Hardware Installation, Optimization/ATP, and FRU 6-71
PRELIMINARY
Table 6-26: Procedure to Test Cabling Calibration using Communication System Analyzer
n ActionStep
2
Enter one or more channel numbers in the Channels box.
NOTE
Multiple channel numbers must be separated by a comma with no spaces (for example: 200,800).
When two or more channel numbers are entered, the cables will be calibrated for each channel.
Interpolation will be accomplished for other channels, as required, for TX calibration.
3 Select TX and RX CABLE CAL, TX CABLE CAL or RX CABLE CAL in the Cable
Calibration picklist.
4 Click OK, and follow the directions displayed for each step. A status report window will be
displayed with the results of the cable calibration.
Calibrate Test Cabling Using
Signal Generator & Spectrum
Analyzer
Refer to Figure 6-21 and follow the procedure in Table 6-27 to calibrate
the TX test cable configuration for all BTSs or the RX ATP test cable
configuration for BTSs with duplexed TX/RX using the signal generator
and spectrum analyzer. Refer to Figure 6-22 and follow the procedure in
Table 6-28 to calibrate the test cable configuration for non–duplexed RX
using the signal generator and spectrum analyzer.
Table 6-27: Procedure to Calibrate TX/Duplexed RX Test Cabling Using
Signal Generator & Spectrum Analyzer
Step Action
1 Connect a short test cable between the spectrum analyzer and the signal generator as shown in
Figure 6-21, detail “A” (top portion of figure).
2 Set signal generator to 0 dBm at the customer frequency of:
869–894 MHz or 1930–1990 MHz
3 Use spectrum analyzer to measure signal generator output (see Figure 6-21, A) and record the value.
4 Connect the spectrum analyzers short cable to point B, (as shown in the lower right portion of the
diagram) to measure cable output at customer frequency of:
869–894 MHz or 1930–1990 MHz
Record the value at point B.
5
Calibration factor = (value measured with detail “A” setup) – (value measured with detail “B” setup)
Example: Cal factor = –1 dBm – (–53.5 dBm) = 52.5 dB
NOTE
The short cable is used for calibration only. It is not part of the final test setup. After calibration is
completed, do not re-arrange any cables. Use the test cable configuration as is to ensure test
procedures use the correct calibration factor.
6