User's Manual

Troubleshooting: Calibration68P64115A18–1
Mar 2003
1X SC 4812T Lite BTS Optimization/ATP Software Release R2.16.1.x
DRAFT
6-11
Calibration Audit Failure
Table 6-12: Troubleshooting Calibration Audit Failure
n Step Action
1 Verify the power meter or communications system analyzer is configured correctly (see the Test
Equipment Set–up section of Chapter 3), and is connected to the proper BTS TX antenna
connector.
2 If a power meter is being used:
2a Re-calibrate the Power Meter and verify it is calibrated correctly with cal factors from the
power sensor (refer to Appendix F).
2b Verify the power sensor is functioning properly by checking it with the 1–mW (0 dBm) Power
Ref signal.
2c Verify communication between the LMF and Power Meter is working by checking that the
meter display is showing RES :
3 Verify that no LPA on the carrier is in alarm state (rapidly flashing red LED).
3a If any are, reset the LPA(s) by pulling the applicable circuit breaker on the DC PDA, and,
after 5 seconds, pushing back in.
4 After calibration, the BLO data must be re-loaded to the BBXs before auditing. Click on the
BBX(s), and in the BTS menu bar select Device > Download >BLO.
Re-try the audit.
5 Verify GPIB adapter is not locked up. Under normal conditions, only 2 green LEDs must be ‘ON’
(Power and Ready). If any other LED is continuously ‘ON’, power-cycle (turn power off and on)
the GPIB Box and retry.
6 If calibration was being performed for the redundant BBX, be sure the Single–Sided BLO
checkbox was not checked in the CDMA Test Parameters test set–up window.
7 If additional items, such as directional couplers or combiners, have been installed in the TX path,
be sure that one of the following has been done:
S Verify BLO checkbox in the CDMA Test Parameters test set–up window is unchecked.
S The additional path losses have been added into each applicable sector using the Util > Edit >
TX Coupler Loss... function.
6