1X SC 4812T Lite BTS Optimization/ATP Software Release R2.16.1.x
DRAFT
F-27
Table F-18: HP 437 Power Meter Calibration Procedure
Step Action
6 Press [ZERO].
Display will show “Zeroing ******.”
Wait for process to complete.
7 Connect the power sensor to the POWER REF output.
8 Turn on the PWR REF by performing the following:
8a Press [SHIFT] then [].
8b Verify that the triangular indicator (below) appears in the display above PWR REF.
9 Perform the following to set the REF CF%:
9a Press ([SHIFT] then [ZERO]) for CAL.
9b Enter the sensors REF CF% from the sensors decal using the arrow keys and press [ENTER].
(The power meter will display ”CAL *****” for a few seconds.)
NOTE
If the REF CAL FACTOR (REF CF) is not shown on the power sensor, assume it to be 100%.
10 Perform the following to set the CAL FAC %:
10a Press [SHIFT] then [FREQ] for CAL FAC.
10b On the sensor’s decal, locate an approximate calibration percentage factor (CF%) at 2 GHz.
10c Enter the sensors calibration % (CF%) using the arrow keys and press [ENTER].
–– When complete, the power meter will typically display 0.05 dBm. (Any reading between
0.00 and 0.10 is normal.)
11 To turn off the PWR REF, perform the following:
11a Press [SHIFT] then [].
11b Disconnect the power sensor from the POWER REF output.