User's Manual

Table Of Contents
Test Set Calibration
68P64115A18–1
Mar 2003
1X SC 4812T Lite BTS Optimization/ATP Software Release R2.16.1.x
DRAFT
3-78
Calibrating Test Cabling using Communications System Analyzer
NOTE
LMF cable calibration cannot be accomplished with an
HP8921A analyzer for 1.9 GHz. A different analyzer type or the
signal generator and spectrum analyzer method (Table 3-27 and
Table 3-28) must be used. Cable calibration values must be
manually entered into the LMF cable loss file if the signal
generator and spectrum analyzer method is used. To use the
HP8921A for manual test cable configuration calibration for 800
MHz BTSs, refer to the Manual Cable Calibration section of
Appendix F.
Prerequisites
S Test equipment is turned on and has warmed up for at least 60
minutes.
S Test equipment has been selected in the LMF (Table 3-23 or
Table 3-24).
S Test equipment has been calibrated and correctly connected for the
type of test cable configuration to be calibrated.
Procedure
Refer to Figure 3-11, Figure 3-12, or Figure 3-13 and follow the
procedure in Table 3-26 to calibrate the test cable configurations.
Table 3-26: Test Cabling Calibration using Comm. System Analyzer
n Step Action
1 Click Util in the BTS menu bar, and select Cable
Calibration... in the pull–down menu. A Cable
Calibration window is displayed.
2
Enter one or more channel numbers in the Channels box.
NOTE
Multiple channel numbers must be separated by a comma
with no spaces (for example,
200,800). When two or
more channel numbers are entered, the cables will be
calibrated for each channel. Interpolation will be
accomplished for other channels, as required, for TX
calibration.
3 Select TX and RX CABLE CAL, TX CABLE CAL or
RX CABLE CAL in the Cable Calibration picklist.
4 Click OK, and follow the directions displayed for each
step. A status report window will be displayed with the
results of the cable calibration.